Skip to content

Datasheet map

The driver follows the FDO2 datasheet. This page says which revision, where to get it, and where each part of the driver is described in it, so you can check any behaviour against the source. Section names and page numbers are the datasheet's own (the printed page numbers).

  • Datasheet: FDO2-G2 Optical Oxygen Sensor Data Sheet, PyroScience Datasheet_FDO2-G2, revision v5 (2026-03). Get it from the vendor page or as a PDF. SHA-256 of the file used: 2644c30ef1bb44384a53777bba9b391d7c91319f4e96f88a94a22bf93455f075.
  • Datasheet: FDO2 Optical Oxygen Sensor Data Sheet (legacy part), PyroScience Datasheet_FDO2, revision v21 (2025-06). Get it from the vendor page or as a PDF. SHA-256 of the file used: 83d026721d08e07e4e82ec268072354f81d07eb92a7fea321f54195e4df6aea5.
  • Product change notification: Low-pass filter and noise erratum, PyroScience PCN-FDO2-G2-001, revision n/a (2026-03). Get it from the vendor page or as a PDF. SHA-256 of the file used: db87761e164dda4256430adec6ade7c5cfd23bf89ea9c28e0f813fc35391adc5.
  • Product change notification: FDO2 replaced by FDO2-G2, PyroScience PCN-FDO2-002, revision n/a (2025-05). Get it from the vendor page or as a PDF. SHA-256 of the file used: c6876c8de992aed439297fefee86ca7d395e3624100ea4bbe6ec0c981af7703f.
  • Quick start guide: Quick Start FDO2-G2 evaluation kit, PyroScience Quick_Start_FDO2-G2, revision v1.00 (n/a). Get it from the vendor page or as a PDF. SHA-256 of the file used: 583d8a7f22a27162533ca44b505a75965d9b1a48fb40e9b39d89fb8f1a67d7a4.
  • Datasheet: FDO2-FTC flow-through cell data sheet (accessory), PyroScience Data Sheet_FDO2_FTC, revision n/a (n/a). Get it from the vendor page or as a PDF. SHA-256 of the file used: 037fea1a8b94016ed95d5be46f2d1ace827d10bb19da605d508df97b6d9f8cd9.
  • Datasheet: FDO2-EXT Modbus and analog extension module data sheet, PyroScience Datasheet_FDO2-EXT, revision v5 (2025-07). Get it from the vendor page or as a PDF. SHA-256 of the file used: bda4536e722f3396de922323287d603f5bb161d1b0263fac9cf80e4c6033eb66.
Topic Datasheet section Page In the driver Explained on
Measuring range, accuracy and warm-up Datasheet_FDO2-G2: 2 Specifications, Analytical Performance p. 5 Config::o2_max_mhpa, Oxygen::warming_up Hardware, Configuration
Supply voltage and current Datasheet_FDO2-G2: 2 Specifications, Interface p. 6 kSupplyMinMv, kSupplyMaxMv, the Power pin hook Hardware
Temperature and pressure limits, humidity Datasheet_FDO2-G2: 2 Specifications, Environment p. 6 Config::temp_min_mc, temp_max_mc, pressure_min_ubar, pressure_max_ubar Hardware, Configuration
Volume fraction from partial pressure Datasheet_FDO2-G2: 2 Specifications, note on % O2 before Environment p. 6 VolumePercentMilli(), MeasureVolumePercentMilli() API guide, Configuration
Mechanics, connector part and weight Datasheet_FDO2-G2: 2 Specifications, Mechanical p. 7 documented in the hardware guide Hardware
Cross-sensitivity and chemical compatibility Datasheet_FDO2-G2: 3 Cross-Sensitivity and Chemical Compatibility p. 8 documented in the hardware guide Hardware, Limitations
UART framing, echo and terminator Datasheet_FDO2-G2: 4.1 General Definitions p. 10 LineAssembler, EncodeCommand(), VerifyEcho() Protocol, Porting
Power-up time Datasheet_FDO2-G2: 4 Communication Interface p. 10 Config::power_up_ms, BeginInitialize() Quickstart, Hardware, Non-blocking use
Numbers in a reply Datasheet_FDO2-G2: 4.1 General Definitions, value formats p. 10 ParseLine(), ParsedLine Protocol
CRC-16/MODBUS suffix Datasheet_FDO2-G2: 4.2 Detecting communication errors and optional CRC p. 11 Crc16Modbus(), CrcMode, CrcActive() Protocol, Settings
#VERS: device, channels, firmware, sensors Datasheet_FDO2-G2: 4.3 Commands, #VERS p. 11 ReadVersion(), Version, CapabilitiesFor() Protocol, API guide
#IDNR: the unique id Datasheet_FDO2-G2: 4.3 Commands, #IDNR p. 11 ReadUniqueId(), Identity::unique_id Protocol, API guide
#MOXY: oxygen, temperature, status Datasheet_FDO2-G2: 4.3 Commands, #MOXY p. 11 MeasureOxygen(), Oxygen Protocol, API guide
#MRAW: phase, signals, pressure, humidity Datasheet_FDO2-G2: 4.3 Commands, #MRAW p. 12 MeasureRaw(), Raw, AmbientLoadMv() Protocol, API guide
Status word bits Datasheet_FDO2-G2: 4.3 Commands, #MOXY, the status word p. 12 Status, Quality, StatusPolicy Status and errors, Configuration
#LOGO: the status LED Datasheet_FDO2-G2: 4.3 Commands, #LOGO p. 13 Identify(), Config::logo_ms Protocol, API guide
#CRCE: the CRC switch Datasheet_FDO2-G2: 4.3 Commands, #CRCE p. 14 SetCrcEnabled(), Config::crc Settings, Protocol
#RDUM and #WRUM: 64 user words Datasheet_FDO2-G2: 4.3 Commands, #RDUM, #WRUM p. 14 ReadUserMemory(), WriteUserMemory() Settings, API guide
#BAUD: the eleven baud rates Datasheet_FDO2-G2: 4.3 Commands, #BAUD p. 14 SetBaudRate(), kBaudRates, Resync() Settings, Porting
#CALO: zero calibration Datasheet_FDO2-G2: 4.3 Commands, #CALO p. 14 CalibrateZero() Calibration
#CAHI: high calibration Datasheet_FDO2-G2: 4.3 Commands, #CAHI p. 15 CalibrateHigh(), CalibrateAir(), AirO2Mhpa() Calibration
#BCST: broadcast mode Datasheet_FDO2-G2: 4.3 Commands, #BCST p. 15 SetBroadcast(), DisableBroadcast(), ProbeBroadcast(), BroadcastPolicy Settings, Non-blocking use
#SETM: measurement modes Datasheet_FDO2-G2: 4.3 Commands, #SETM p. 15 SetMeasurementMode(), Capabilities::set_mode Settings
Measurement duration Datasheet_FDO2-G2: 4.3 Commands, #SETM, modes p. 16 Config::measurement_ms Configuration, Non-blocking use
Flash endurance and interrupted writes Datasheet_FDO2-G2: 4.3 Commands, WARNING p. 16 Config::allow_persistent_writes, max_flash_writes, Error::StateUnknown Calibration, Contract, Limitations
#ERRO codes Datasheet_FDO2-G2: 4.4 Error Codes, Table 1 p. 17 Error, DeviceErrorCode() Status and errors, Contract
Firmware revision history of the G2 Datasheet_FDO2-G2: 5 Released Firmware Revisions Changes p. 18 CapabilitiesFor(), Identity::noise_erratum Protocol, Limitations
Errata: noise at two temperatures Datasheet_FDO2-G2: 6 ERRATA, 6.1 Increased Noise at Specific Temperatures p. 19 HasNoiseErratum(), DisableBroadcast() Limitations, Settings, Troubleshooting
Connector, pins and signal levels Datasheet_FDO2-G2: 7 Mechanical Dimensions and Electrical Interface, drawing and pin table p. 20 CtrlPin, GpioSignal, kLogicLevelMv, kMaxLogicLevelMv Hardware, Quickstart, Porting
Mechanical drawing Datasheet_FDO2-G2: 7 Mechanical Dimensions and Electrical Interface, drawings p. 20 documented in the hardware guide Hardware
Warnings: no medical or safety use Datasheet_FDO2-G2: 8 Warnings p. 21 documented in the limitations Limitations
Supply below 3.3 V and readings Datasheet_FDO2: 2 Specifications, Interface, WARNING p. 6 documented in the guides; no driver check (the module has no supply reading) Hardware, Limitations
Firmware revision history Datasheet_FDO2: 5 Released Firmware Revisions Changes p. 17 CapabilitiesFor(), kFirmware* Protocol, Limitations
UART levels: 3.0 V, 3.3 V maximum on RXD Datasheet_FDO2: 6.2 Voltage Tolerance of UART RX Pin (ERRATA) p. 19 kMaxLogicLevelMv, port static_asserts in the examples Hardware, Porting, Troubleshooting
Errata: broadcast send and receive conflict (firmware 3.51) Datasheet_FDO2: 6.1 Firmware Bug Broadcast Send and Receive Conflict (FW 3.51) p. 19 DisableBroadcast() (waits for a frame boundary), CapabilitiesFor() Settings, Limitations, Troubleshooting
Errata: error handling of the internal temperature sensor Datasheet_FDO2: 6.3 Error Handling of Internal Temperature Sensor p. 20 Capabilities::temp_flag_reliable, Status::TemperatureSensorFailed() Status and errors, Limitations
Low-pass filter change and the noise erratum PCN-FDO2-G2-001: Product change notification, text p. 1 Identity::noise_erratum Limitations
Replacement of the FDO2 by the FDO2-G2 PCN-FDO2-002: Product change notification, text p. 1 Family Hardware, Limitations
Start-up flashes of the status LED Quick_Start_FDO2-G2: Quick start, connecting the module p. 2 documented in the quickstart Quickstart
Flow-through cell accessory Data Sheet_FDO2_FTC: Flow-through cell data sheet, whole document p. 1 none (an accessory) Hardware
Modbus extension module (a different interface) Datasheet_FDO2-EXT: 1 INTRODUCTION p. 1 none (out of scope) Limitations

When the datasheet and the driver differ

The driver is written against the revision above. If you have a newer revision, compare its revision history with the topics above; a change to a pin, a register, a timing or a value the driver uses is a bug report for this repository. Where the driver deliberately goes beyond the datasheet, the Contract says so.

The repository never contains the PDF itself: vendors own their documents. Download it from the vendor, and compare its SHA-256 with the one above to confirm you have the same file.