Datasheet map¶
The driver follows the FDO2 datasheet. This page says which revision, where to get it, and where each part of the driver is described in it, so you can check any behaviour against the source. Section names and page numbers are the datasheet's own (the printed page numbers).
- Datasheet: FDO2-G2 Optical Oxygen Sensor Data Sheet, PyroScience Datasheet_FDO2-G2, revision v5 (2026-03). Get it
from the vendor page or as a
PDF.
SHA-256 of the file used:
2644c30ef1bb44384a53777bba9b391d7c91319f4e96f88a94a22bf93455f075. - Datasheet: FDO2 Optical Oxygen Sensor Data Sheet (legacy part), PyroScience Datasheet_FDO2, revision v21
(2025-06). Get it from the vendor page or as a
PDF.
SHA-256 of the file used:
83d026721d08e07e4e82ec268072354f81d07eb92a7fea321f54195e4df6aea5. - Product change notification: Low-pass filter and noise erratum, PyroScience PCN-FDO2-G2-001, revision n/a
(2026-03). Get it from the vendor page or as a
PDF.
SHA-256 of the file used:
db87761e164dda4256430adec6ade7c5cfd23bf89ea9c28e0f813fc35391adc5. - Product change notification: FDO2 replaced by FDO2-G2, PyroScience PCN-FDO2-002, revision n/a (2025-05). Get it
from the vendor page or as a
PDF.
SHA-256 of the file used:
c6876c8de992aed439297fefee86ca7d395e3624100ea4bbe6ec0c981af7703f. - Quick start guide: Quick Start FDO2-G2 evaluation kit, PyroScience Quick_Start_FDO2-G2, revision v1.00 (n/a). Get
it from the vendor page or as a
PDF.
SHA-256 of the file used:
583d8a7f22a27162533ca44b505a75965d9b1a48fb40e9b39d89fb8f1a67d7a4. - Datasheet: FDO2-FTC flow-through cell data sheet (accessory), PyroScience Data Sheet_FDO2_FTC, revision n/a (n/a).
Get it from the vendor page or as a
PDF.
SHA-256 of the file used:
037fea1a8b94016ed95d5be46f2d1ace827d10bb19da605d508df97b6d9f8cd9. - Datasheet: FDO2-EXT Modbus and analog extension module data sheet, PyroScience Datasheet_FDO2-EXT, revision v5
(2025-07). Get it from the vendor
page or as a
PDF.
SHA-256 of the file used:
bda4536e722f3396de922323287d603f5bb161d1b0263fac9cf80e4c6033eb66.
| Topic | Datasheet section | Page | In the driver | Explained on |
|---|---|---|---|---|
| Measuring range, accuracy and warm-up | Datasheet_FDO2-G2: 2 Specifications, Analytical Performance | p. 5 | Config::o2_max_mhpa, Oxygen::warming_up |
Hardware, Configuration |
| Supply voltage and current | Datasheet_FDO2-G2: 2 Specifications, Interface | p. 6 | kSupplyMinMv, kSupplyMaxMv, the Power pin hook |
Hardware |
| Temperature and pressure limits, humidity | Datasheet_FDO2-G2: 2 Specifications, Environment | p. 6 | Config::temp_min_mc, temp_max_mc, pressure_min_ubar, pressure_max_ubar |
Hardware, Configuration |
| Volume fraction from partial pressure | Datasheet_FDO2-G2: 2 Specifications, note on % O2 before Environment | p. 6 | VolumePercentMilli(), MeasureVolumePercentMilli() |
API guide, Configuration |
| Mechanics, connector part and weight | Datasheet_FDO2-G2: 2 Specifications, Mechanical | p. 7 | documented in the hardware guide | Hardware |
| Cross-sensitivity and chemical compatibility | Datasheet_FDO2-G2: 3 Cross-Sensitivity and Chemical Compatibility | p. 8 | documented in the hardware guide | Hardware, Limitations |
| UART framing, echo and terminator | Datasheet_FDO2-G2: 4.1 General Definitions | p. 10 | LineAssembler, EncodeCommand(), VerifyEcho() |
Protocol, Porting |
| Power-up time | Datasheet_FDO2-G2: 4 Communication Interface | p. 10 | Config::power_up_ms, BeginInitialize() |
Quickstart, Hardware, Non-blocking use |
| Numbers in a reply | Datasheet_FDO2-G2: 4.1 General Definitions, value formats | p. 10 | ParseLine(), ParsedLine |
Protocol |
| CRC-16/MODBUS suffix | Datasheet_FDO2-G2: 4.2 Detecting communication errors and optional CRC | p. 11 | Crc16Modbus(), CrcMode, CrcActive() |
Protocol, Settings |
| #VERS: device, channels, firmware, sensors | Datasheet_FDO2-G2: 4.3 Commands, #VERS | p. 11 | ReadVersion(), Version, CapabilitiesFor() |
Protocol, API guide |
| #IDNR: the unique id | Datasheet_FDO2-G2: 4.3 Commands, #IDNR | p. 11 | ReadUniqueId(), Identity::unique_id |
Protocol, API guide |
| #MOXY: oxygen, temperature, status | Datasheet_FDO2-G2: 4.3 Commands, #MOXY | p. 11 | MeasureOxygen(), Oxygen |
Protocol, API guide |
| #MRAW: phase, signals, pressure, humidity | Datasheet_FDO2-G2: 4.3 Commands, #MRAW | p. 12 | MeasureRaw(), Raw, AmbientLoadMv() |
Protocol, API guide |
| Status word bits | Datasheet_FDO2-G2: 4.3 Commands, #MOXY, the status word | p. 12 | Status, Quality, StatusPolicy |
Status and errors, Configuration |
| #LOGO: the status LED | Datasheet_FDO2-G2: 4.3 Commands, #LOGO | p. 13 | Identify(), Config::logo_ms |
Protocol, API guide |
| #CRCE: the CRC switch | Datasheet_FDO2-G2: 4.3 Commands, #CRCE | p. 14 | SetCrcEnabled(), Config::crc |
Settings, Protocol |
| #RDUM and #WRUM: 64 user words | Datasheet_FDO2-G2: 4.3 Commands, #RDUM, #WRUM | p. 14 | ReadUserMemory(), WriteUserMemory() |
Settings, API guide |
| #BAUD: the eleven baud rates | Datasheet_FDO2-G2: 4.3 Commands, #BAUD | p. 14 | SetBaudRate(), kBaudRates, Resync() |
Settings, Porting |
| #CALO: zero calibration | Datasheet_FDO2-G2: 4.3 Commands, #CALO | p. 14 | CalibrateZero() |
Calibration |
| #CAHI: high calibration | Datasheet_FDO2-G2: 4.3 Commands, #CAHI | p. 15 | CalibrateHigh(), CalibrateAir(), AirO2Mhpa() |
Calibration |
| #BCST: broadcast mode | Datasheet_FDO2-G2: 4.3 Commands, #BCST | p. 15 | SetBroadcast(), DisableBroadcast(), ProbeBroadcast(), BroadcastPolicy |
Settings, Non-blocking use |
| #SETM: measurement modes | Datasheet_FDO2-G2: 4.3 Commands, #SETM | p. 15 | SetMeasurementMode(), Capabilities::set_mode |
Settings |
| Measurement duration | Datasheet_FDO2-G2: 4.3 Commands, #SETM, modes | p. 16 | Config::measurement_ms |
Configuration, Non-blocking use |
| Flash endurance and interrupted writes | Datasheet_FDO2-G2: 4.3 Commands, WARNING | p. 16 | Config::allow_persistent_writes, max_flash_writes, Error::StateUnknown |
Calibration, Contract, Limitations |
| #ERRO codes | Datasheet_FDO2-G2: 4.4 Error Codes, Table 1 | p. 17 | Error, DeviceErrorCode() |
Status and errors, Contract |
| Firmware revision history of the G2 | Datasheet_FDO2-G2: 5 Released Firmware Revisions Changes | p. 18 | CapabilitiesFor(), Identity::noise_erratum |
Protocol, Limitations |
| Errata: noise at two temperatures | Datasheet_FDO2-G2: 6 ERRATA, 6.1 Increased Noise at Specific Temperatures | p. 19 | HasNoiseErratum(), DisableBroadcast() |
Limitations, Settings, Troubleshooting |
| Connector, pins and signal levels | Datasheet_FDO2-G2: 7 Mechanical Dimensions and Electrical Interface, drawing and pin table | p. 20 | CtrlPin, GpioSignal, kLogicLevelMv, kMaxLogicLevelMv |
Hardware, Quickstart, Porting |
| Mechanical drawing | Datasheet_FDO2-G2: 7 Mechanical Dimensions and Electrical Interface, drawings | p. 20 | documented in the hardware guide | Hardware |
| Warnings: no medical or safety use | Datasheet_FDO2-G2: 8 Warnings | p. 21 | documented in the limitations | Limitations |
| Supply below 3.3 V and readings | Datasheet_FDO2: 2 Specifications, Interface, WARNING | p. 6 | documented in the guides; no driver check (the module has no supply reading) | Hardware, Limitations |
| Firmware revision history | Datasheet_FDO2: 5 Released Firmware Revisions Changes | p. 17 | CapabilitiesFor(), kFirmware* |
Protocol, Limitations |
| UART levels: 3.0 V, 3.3 V maximum on RXD | Datasheet_FDO2: 6.2 Voltage Tolerance of UART RX Pin (ERRATA) | p. 19 | kMaxLogicLevelMv, port static_asserts in the examples |
Hardware, Porting, Troubleshooting |
| Errata: broadcast send and receive conflict (firmware 3.51) | Datasheet_FDO2: 6.1 Firmware Bug Broadcast Send and Receive Conflict (FW 3.51) | p. 19 | DisableBroadcast() (waits for a frame boundary), CapabilitiesFor() |
Settings, Limitations, Troubleshooting |
| Errata: error handling of the internal temperature sensor | Datasheet_FDO2: 6.3 Error Handling of Internal Temperature Sensor | p. 20 | Capabilities::temp_flag_reliable, Status::TemperatureSensorFailed() |
Status and errors, Limitations |
| Low-pass filter change and the noise erratum | PCN-FDO2-G2-001: Product change notification, text | p. 1 | Identity::noise_erratum |
Limitations |
| Replacement of the FDO2 by the FDO2-G2 | PCN-FDO2-002: Product change notification, text | p. 1 | Family |
Hardware, Limitations |
| Start-up flashes of the status LED | Quick_Start_FDO2-G2: Quick start, connecting the module | p. 2 | documented in the quickstart | Quickstart |
| Flow-through cell accessory | Data Sheet_FDO2_FTC: Flow-through cell data sheet, whole document | p. 1 | none (an accessory) | Hardware |
| Modbus extension module (a different interface) | Datasheet_FDO2-EXT: 1 INTRODUCTION | p. 1 | none (out of scope) | Limitations |
When the datasheet and the driver differ¶
The driver is written against the revision above. If you have a newer revision, compare its revision history with the topics above; a change to a pin, a register, a timing or a value the driver uses is a bug report for this repository. Where the driver deliberately goes beyond the datasheet, the Contract says so.
The repository never contains the PDF itself: vendors own their documents. Download it from the vendor, and compare its SHA-256 with the one above to confirm you have the same file.