What to check when a call fails, by the error it returns.

Datasheet sections this page relies on
  • Errata: noise at two temperatures: PyroScience Datasheet_FDO2-G2 v5, 6 ERRATA, 6.1 Increased Noise at Specific Temperatures, p. 19
  • UART levels: 3.0 V, 3.3 V maximum on RXD: PyroScience Datasheet_FDO2 v21, 6.2 Voltage Tolerance of UART RX Pin (ERRATA), p. 19
  • Errata: broadcast send and receive conflict (firmware 3.51): PyroScience Datasheet_FDO2 v21, 6.1 Firmware Bug Broadcast Send and Receive Conflict (FW 3.51), p. 19

All of them, with the driver API for each: Datasheet map.

EnsureInitialized returns Bus

A port hook reported a failure.

  1. Check that your port's EnsureInitialized() returns true on its own (log it).
  2. Check the UART pins in your port against the wiring.
  3. Check that Write() returns true and sends all the bytes.

EnsureInitialized returns Timeout

The module did not answer #VERS.

Symptom Likely cause Fix
Nothing on TXD No power, or below 3.3 V Check VCC (3.3 V to 5.0 V) and GND
Nothing on TXD TX and RX swapped Cross them: MCU TX to RXD, MCU RX to TXD
Garbage on TXD The module is at another baud rate Power-cycle it: it starts at 19200
The module streams lines A broadcast is on See below
Answers only after a wait A short power_up_ms Raise Config::power_up_ms

EnsureInitialized returns DeviceIdMismatch

#VERS answered, but not as an FDO2 with an oxygen sensor: the device id is not 8 or the sensor bits lack oxygen. Check that the line is not shared with another device.

EnsureInitialized returns BroadcastActive

The module streams #MRAW lines (#BCST is on) and Config::broadcast is Refuse. Set it to Stream to read the frames, or to Disable to turn the stream off (a flash write: it needs allow_persistent_writes). A slow stream is seen only when broadcast_listen_ms is at least its interval (Settings).

Calls return Crc now and then

Noise on the line. Shorten the wires, keep them away from the supply and check the ground. The driver already retried a read; GetCounters().crc_errors shows how often it happens.

Calls return InvalidReply or EchoMismatch

Bytes that are not part of the reply, or a wrong baud rate. Check that FlushInput() really flushes, that the receive buffer holds 800 bytes and that nothing else writes to the line.

Calls return SensorFault

The status word is not acceptable under the status policy. Status and errors lists the bits. With the sensor in air and a signal too low, check that the membrane is clean and uncovered.

Calls return StateUnknown

A flash write or a baud change ended with no echo. Call Resync(): it tries the configured rate, the new rate and 19200, each with and without the checksum, and reads #VERS. Then read the setting back before you write it again.

Calls return NotPermitted

A flash write needs allow_persistent_writes and room in max_flash_writes (Calibration).

Calls return Unsupported

The firmware answered #ERRO -26 (it does not know the command), or your port has no hook the call needs (SetBaudRate). GetIdentity().caps shows what the firmware offers.

Readings look wrong

  • Within the first three minutes after power-up the accuracy is reduced (warming_up).
  • A supply below 3.3 V can make the readings wrong; the module has no supply reading.
  • A calibration at the wrong pressure, humidity or temperature moves every later reading (Calibration).