Calibration

How to calibrate the module in air or with two points, and why every calibration call needs your consent first.

Datasheet sections this page relies on

From PyroScience Datasheet_FDO2-G2 v5:

  • CALO: zero calibration: 4.3 Commands, #CALO, p. 14

  • CAHI: high calibration: 4.3 Commands, #CAHI, p. 15

  • Flash endurance and interrupted writes: 4.3 Commands, WARNING, p. 16

All of them, with the driver API for each: Datasheet map.

#CALO and #CAHI write the module's flash, and so do #CRCE, #WRUM, #BCST and #SETM. The flash survives only a limited number of writes (the module's own limit, about 20000 cycles in all). So the driver sends these commands only when you ask for it in the Config:

configuration.cpp
fdo2::Config WritableConfig() {
  fdo2::Config c{};
  c.allow_persistent_writes = true; // every flash-writing call needs this
  c.max_flash_writes = 4; // and is counted against this budget (about 20000 cycles in all)
  return c;
}
  • allow_persistent_writes is false by default. A flash-writing call returns NotPermitted and sends nothing.
  • max_flash_writes (1 to 64, default 8) is the number of such calls the driver sends in its lifetime. FlashWritesUsed() shows the count. A write counts when it is sent, even if the port fails, because the module may have received it.
  • A flash write is never repeated. If the reply is lost, the call returns the original error (for example Timeout) and the state becomes StateUnknown; call Resync() and read the module back before you try again.

Vendor warnings

The datasheet (section 4.3, Commands, and the warning that follows it) says, and the driver follows:

  • Flash endurance is about 20,000 flash cycles in all, triggered by #CALO, #CAHI, #CRCE, #SETM and #BCST (and by #WRUM, which the datasheet describes as one cycle per call). Use them only when needed. max_flash_writes is a guard against a loop in your code, not the vendor figure: its limit of 64 per driver lifetime is far below 20,000.
  • A power loss during a flash cycle can corrupt the flash. Keep the supply stable while one runs. The driver refuses Deinitialize() with Busy while a write is pending, and PowerCycle() waits for it to end (Settings).
  • Always evaluate the reply. A write without its echo may have corrupted the flash: the driver reports the original error and the state StateUnknown.
  • #CALO and #CAHI replace the factory calibration. The datasheet asks for at least 5 minutes of steady measuring in the reference medium before the command, and the command takes up to 10 seconds.

The guard before every flash write

A wrong calibration is stored in the module for good, so the driver checks before it sends #CALO or #CAHI:

  1. It reads #VERS and compares it with the identity found at start-up. A different module on the line, or a module that has reset without the driver seeing it (a power glitch, for example), ends in StateUnknown or DeviceIdMismatch and nothing is written.
  2. It reads #MOXY. The status must be acceptable, and the sensor must have been powered for the warm-up time of 3 minutes (datasheet section 2): earlier the call returns NotReady. The measured oxygen partial pressure must lie within Config::cal_window_pct percent (default 20, at least cal_window_min_mhpa, default 5000 milli-hPa) of the value you are about to store. A sensor still in air cannot be calibrated as nitrogen, and the reverse: the call returns OutOfRange and nothing is written.
  3. Only then does it send the write, and it counts it.

The guard reads #MOXY but not #VERS before a calibration, and its warm-up test counts the time since the driver powered the module, using the host's clock (the datasheet gives no uptime value to read). A module that was swapped or reset after start-up is therefore seen only through the plausibility window and the status word.

The other flash commands (#CRCE, #WRUM, #BCST, #SETM) get the identity check (1). The one exception is #BCST 0 while the module streams, which is sent as it is because no reply can be read in a stream. CalibrateHigh() accepts 10 to 2000 hPa (10000 to 2000000 in 0.001 hPa): a standard outside the range of a real calibration gas returns InvalidArgument.

Before you calibrate

  • Keep the sensor in the reference medium until the readings are steady. The datasheet asks for five minutes. steady_s is the time you have waited; a value below Config::min_steady_s (default 300) returns InvalidArgument and sends nothing.
  • The status must be good. CalibrateAir() returns SensorFault otherwise.
  • The sensor must be warmed up (3 minutes after power-up), or the call returns NotReady.
  • Calibration changes the module's stored calibration. Not verified on hardware: nobody has run these calls against a module yet.

One point in air

calibration.cpp
// One high point in ambient air; the sensor has been in it for at least five minutes.
fdo2::Error CalibrateInAir(Sensor& sensor) {
  // Reads the module's pressure, takes 20.95 % oxygen (dry air) and sends #CAHI.
  return sensor.CalibrateAir(300).error; // NotPermitted unless allowed in the Config
}

CalibrateAir() reads #MRAW for the pressure, computes the oxygen partial pressure of dry air at that pressure (20.95 percent by default, kDryAirO2Ppm) and sends #CAHI. It is correct only when the sensor sees dry air at the pressure of the module's own sensor, which sees the air through the venting capillary at the back of the housing (datasheet section 2): keep the capillary open to the same air as the membrane. Humid air has a lower oxygen partial pressure; the datasheet does not give a correction, so the driver applies none. You can see the value before anything is written:

calibration.cpp
// The value #CAHI will receive, before anything is written.
int32_t AirValue(int32_t pressure_ubar) {
  const auto v = fdo2::AirO2Mhpa(pressure_ubar); // 1013000 ubar -> 212224 (10^-3 hPa)
  return v ? v.value : 0;
}

Two points

calibration.cpp
// Zero in nitrogen, then a known partial pressure; each step needs its own steady time.
fdo2::Error CalibrateTwoPoint(Sensor& sensor, int32_t high_po2_mhpa) {
  if (const auto r = sensor.CalibrateZero(300); !r) {
    return r.error; // CalibrationRejected: #ERRO -60, the value is outside the expected range
  }
  return sensor.CalibrateHigh(high_po2_mhpa, 300).error;
}

CalibrateZero() sends #CALO with the sensor in an oxygen-free medium. CalibrateHigh() sends #CAHI with the oxygen partial pressure of your standard in 0.001 hPa. A call allows Config::calibration_ms (10 s by default) for the reply.

Errors

Error Meaning
NotPermitted No consent, or the write budget is used
InvalidArgument steady_s too short, or the value is out of range (10 to 2000 hPa)
NotReady The sensor is still warming up (3 minutes after power-up)
OutOfRange The measured value is far from the calibration value (the guard)
DeviceIdMismatch The module on the line is not the one found at start-up (the guard)
CalibrationRejected #ERRO -60: the module judged the value outside its expected range (firmware 3.51 and later)
SensorFault The status word was not good (CalibrateAir())
StateUnknown The reply was lost; Resync()