Datasheet map¶
The driver follows the Sensirion documents listed below. This page says which revision, where to get it, where each part of the driver is described in it, and which datasheet features the driver covers, so you can check any behaviour against the source. Section names and page numbers are the documents' own (the printed page numbers of the SFM4300 unless a row names another document).
- Datasheet: SFM3000 digital mass flow meter, Sensirion AG SFM3000, revision 2.2 (July 2016). Get it from the
vendor page.
SHA-256 of the file used:
2e29868908c3f5313c9d3381cbdc74e9dbdab35f81116d147f4da626110c4654. - Datasheet: SFM3003 digital mass flow meter, Sensirion AG SFM3003 D1, revision 1.1 (June 2024). Get it from the
vendor page.
SHA-256 of the file used:
735ab7a5b8df25991788bb9e35b16fc77402096c8060f791c9db2983f6faac89. - Datasheet: SFM3003-300-CET digital mass flow meter, Sensirion AG SFM3003-300-CET, revision 1.0 (April 2022). Get
it from the vendor
page.
SHA-256 of the file used:
1202f25aa86bb544ecc5e49a245d3f1fefe299b167f0902cb5e72d5fc6676434. - Datasheet: SFM3013 digital mass flow meter, Sensirion AG SFM3013 D3, revision 1.1 (June 2024). Get it from the
vendor page.
SHA-256 of the file used:
a0ee3acf1bbd22e8e9f988ae7a0f28099beeab80d2775f80ff60909e8724003b. - Datasheet: SFM3019 digital mass flow meter, Sensirion AG SFM3019 D1, revision 1.3 (not printed (vendor page:
08/2024)). Get it from the vendor
page.
SHA-256 of the file used:
d9c1b2e92f113a3f75a1b3dc7a3b5f552b3963a748ee919191a209bc6214d510. - Datasheet: SFM3020 analog mass flow meter (reviewed, not supported), Sensirion AG SFM3020, revision 1.2 (not
printed). Get it from the vendor
page.
SHA-256 of the file used:
13d7ade4def1e7ff884b54784b97001cdc26c8f5402e25fbc3d5ecbb34077451. - Datasheet: SFM3100 analog mass flow meter (reviewed, not supported), Sensirion AG SFM3100, revision 1.5 (not
printed (vendor page: 08/2026)). Get it from the vendor
page.
SHA-256 of the file used:
6a9b7f4e28e003419fe0d8c62fbad4017d1d57f1caaff87bd97916f7d0b8680f. - Datasheet: SFM3119 digital mass flow meter, Sensirion AG SFM3119 D1, revision 1.3 (September 2023). Get it from
the vendor page.
SHA-256 of the file used:
0ba4486614528477e0fd7b19ed872375eeeb2039a8ab29f734a57c24a1c2698d. - Datasheet: SFM3200 digital mass flow meter, Sensirion AG SFM3200, revision 1.1 (May 2017). Get it from the vendor
page or as a
PDF.
SHA-256 of the file used:
02bee98ec63a88656dfbb7a16c7e036ea1ea1025467a7b98c400703b78f67ca2. - Datasheet: SFM3200-AW digital mass flow meter, Sensirion AG SFM3200-AW, revision 1.1 (March 2022). Get it from the
vendor page.
SHA-256 of the file used:
de9cf028a1fe8a5caed50644aa5ed279c8b26cca447908afcb9450232e91aa4c. - Datasheet: SFM3300-D and SFM3300-AW digital mass flow meters, Sensirion AG SFM3300, revision 1.4 (March 2022). Get
it from the vendor
page.
SHA-256 of the file used:
f7c20973d659290345ac1ac725da5c46c57924b255d64d14754aaac3c8bbd1b8. - Datasheet: SFM3304-xxx-D digital mass flow meter, Sensirion AG SFM3304, revision 1.3 (not printed (vendor page:
09/2026)). Get it from the vendor
page or as a
PDF.
SHA-256 of the file used:
5cb6f4ac1951e6e64ee85e1202182bd998d9734bd185ccb97eced7cd4130818c. - Application note: Transition guide SFM3300-250-D to SFM3304-250-D, Sensirion AG SFM3300-SFM3304 transition,
revision 1 (not printed (vendor page: 10/2026)). Get it from the vendor
page.
SHA-256 of the file used:
79ed3307bd97b1426d759a4b1ab1950857f907c0455d68a0c638707371459a44. - Datasheet: SFM3400-AW and SFM3400-D digital mass flow meters, Sensirion AG SFM3400, revision 1.5 (January 2024).
Get it from the vendor page.
SHA-256 of the file used:
5213e5e768f4ebd93593241d346c091ca382196d2062b0300ea158ae85071d23. - Datasheet: SFM3505 digital mass flow meter, Sensirion AG SFM3505 D3, revision 1.0 (October 2025). Get it from the
vendor page or as a
PDF.
SHA-256 of the file used:
8d378b626e7ad4bf476d850ac3787132a3a97d6b3de8bc2d955ff08cd3ace867. - Datasheet: SFM4100 series digital mass flow meters, Sensirion AG SFM4100, revision 1.4 (January 2024). Get it from
the vendor page or as a
PDF.
SHA-256 of the file used:
88ec21b5ddc862a7367fd2c70a14601987d4f7785c2bf252eede4e954143db59. - Application note: CRC calculation for the SFM4100, Sensirion AG SFM4100 CRC, revision 1.0 (June 2009). Get it from
the vendor
page or as
a PDF.
SHA-256 of the file used:
cbc9c48f190ea7daffd5a9f048555767b5eb83e799438387654ec448c8147787. - Application note: SFM4100 sample code, Sensirion AG SFM4100 sample code, revision 1.0 (July 2009). Get it from the
vendor
page.
SHA-256 of the file used:
ec8370b9b186167eb970a5004811e8ae3ab2c082caff03d16898bf1ea93b6e91. - Datasheet: SFM4200 digital mass flow meter, Sensirion AG SFM4200 D1, revision 1.2 (August 2018). Get it from the
vendor page.
SHA-256 of the file used:
4e72f7be81414f7971691ce595e01fa142d90a2a9fbfe2238caa65d43f884352. - Datasheet: SFM4300 series digital mass flow meters, Sensirion AG SFM4300 D1, revision 1.0 (April 2022). Get it
from the vendor page or as a
PDF.
SHA-256 of the file used:
51d3c45adeef518764a413d46624dffc5c6ce7388b3456c8998342c6f33684ec. - Datasheet: SFC5300 and SFM5300 (2016; reviewed, not supported), Sensirion AG SFM5300 (2016), revision not printed
(2016). Get it from the vendor
page.
SHA-256 of the file used:
73f8e01e632fc277fe20335117b6cc8e2694a61b0b2d2bd49a5a8c6a15e404ef. - Datasheet: SFC5300/SFM5300 and SFC5330/SFM5330 (reviewed, not supported), Sensirion AG SFx53x0, revision 05/2023
(May 2023). Get it from the vendor
page.
SHA-256 of the file used:
a74155724e53b55c6696cfbd187fc4c78137d5ad5a069e08bde1240cc28dfdc0. - Datasheet: SFC5400 and SFM5400 (reviewed, not supported), Sensirion AG SFx5400, revision 05/2023 (May 2023). Get
it from the vendor page.
SHA-256 of the file used:
b0072f8b78af49be3abca9d42ca5960be4676f4aeb64fada8c46c853bb3b4907. - Datasheet: SFC5460 and SFM5460 (reviewed, not supported), Sensirion AG SFx5460, revision 05/2023 (May 2023). Get
it from the vendor page.
SHA-256 of the file used:
74e6d4a3568a2bc9fb3bfe5758f12fefb46e340c7108578fa1c2287d5a9b99c3. - Datasheet: SFC5500 and SFM5500 (reviewed, not supported), Sensirion AG SFx5500, revision 02/2024 (February 2024).
Get it from the vendor page.
SHA-256 of the file used:
a41a5c4910e15b458642da87a3bfd1159b3bb89ee44f221a512ac88c87ac9c71. - Reference manual: SFC5xxx SHDLC interface reference (reviewed, not supported), Sensirion AG SFC5xxx SHDLC,
revision 1.9 (October 2015). Get it from the vendor
page.
SHA-256 of the file used:
e32f664a72a8d9cb008fa820209ae5d966ab08ab18cc6fda23230a0379852f94. - Datasheet: SFC6000D and SFM6000D digital mass flow meters, Sensirion AG SFx6000D D1, revision 1.1 (May 2023). Get
it from the vendor page.
SHA-256 of the file used:
4e633254f7b26694059da5785d90946dfa9a424ae159f7206858998f9a54b870. - Reference manual: SFC6xxx/SFM6xxx I2C interface reference (preliminary), Sensirion AG SFx6000 I2C, revision 1.1
(not printed (vendor page: 07/2023)). Get it from the vendor
page or as a
PDF.
SHA-256 of the file used:
6a6e9f48ec98cbeba48a5ba9a610e2d9d89d09b1faee55c23fb20c7b0defa051. - Reference manual: SFC6xxx/SFM6xxx RS485 SHDLC guide (reviewed, not supported), Sensirion AG SFx6000 SHDLC,
revision 1.1 (not printed). Get it from the vendor
page.
SHA-256 of the file used:
13d23c87e3cc79363bf7101ce0c4379485c499f94a3a11e12e0aa3bd18b8f286. - Application note: SFM-05 I2C functional description for the SFM3x00 series, Sensirion AG SFM-05, revision not
printed (vendor page: 10/2026) (not printed). Get it from the vendor
page or as a
PDF.
SHA-256 of the file used:
f45cd8d3d8ef50655dd70139b6020a0a85c1dc3c1a1aa09b446c4308e272730d. - Application note: SFM-04 CRC checksum, Sensirion AG SFM-04 D1, revision not printed (not printed (vendor page:
01/2020)). Get it from the vendor
page.
SHA-256 of the file used:
066f8e4631cbdd8ef87dd1f14d4a0abb1effe42b85d050bc94a6f999afb0b557. - Application note: SFM-01 engineering guidelines for mass flow meters, Sensirion AG SFM-01, revision not printed
(not printed). Get it from the vendor
page.
SHA-256 of the file used:
f3b68ecd39b3feeaf1e678183e1576334a117de4259d43b6938901c95267fb8d.
| Topic | Datasheet section | Page | In the driver | Explained on |
|---|---|---|---|---|
| Sf05 reset, restart and first read | SFM3200: Digital interface (soft reset 80 ms, restart before read) | p. 5 | Config::restart_before_read, Config::consecutive_error_limit |
Architecture, Configuration, Hardware |
| Gen2 start with a filter argument and its status layout | SFM3304: Digital interface description (start 0x3603 with filter, status word) | p. 11 | Family::Sf06Gen2, Config::filter_enabled, Config::filter |
API guide, Architecture |
| Dual-gas frame of 24 bits | SFM3505 D3: Digital interface description (24-bit frame, commands) | p. 10 | Family::Sf06Dual, Sample::flow2_ml_per_min, Sample::has_flow2 |
API guide, Architecture, Configuration |
| Sf04 trigger bytes and the hold-master read | SFM4100: Digital interface (trigger, clock stretching, address per gas) | p. 4 | Family::Sf04, ReadStretch(), kClockStretch |
Porting, Architecture, API guide, Hardware |
| Sf04 CRC | SFM4100 CRC: CRC calculation | p. 2 | reg::Crc8() with init 0x00 |
Contract, Testing and CI |
| Timing (reset, first sample, wake-up, stop) | SFM4300 D1: 2.4 Timing specification | p. 5 | Config::timing, TimingPolicy, Wait, DelayUs() |
Hardware, Porting, Architecture |
| Electrical specification and limits | SFM4300 D1: 2.3 Electrical specification; 2.7 Absolute maximum ratings | p. 5 | - | Hardware |
| Connector and pins | SFM4300 D1: 2.8 Electrical connection, Table 1 | p. 6 | - | Hardware |
| IRQn and start-up behaviour | SFM4300 D1: 3.2 Sensor start-up and warm-up; 3.3 IRQn | p. 8 | Sample::settled, Config::require_settled, Error::DeviceReset |
Hardware, Architecture, Troubleshooting |
| I2C address and bus | SFM4300 D1: 4 Digital interface description; 4.1 I2C address | p. 9 | I2cInterface, I2cPort, Config::address, PartInfo::address |
Hardware, Porting, Architecture |
| Command framing and read framing | SFM4300 D1: 4.2 I2C sequences | p. 9 | Write(), Read(), detail::ParseWords() |
Porting, Architecture, Contract |
| Start measurement, results and the NACK of a read without data | SFM4300 D1: 4.3.1 Start continuous measurement | p. 10 | StartMeasurement(), ReadSample(), Error::NoData, Error::DeviceReset |
Quickstart, API guide, Architecture, Troubleshooting |
| Status word | SFM4300 D1: 4.3.2 Read measurement results, Table 5 | p. 11 | Sample::status, Error::StatusMismatch |
API guide, Contract, Troubleshooting |
| Concentration update | SFM4300 D1: 4.3.3 Update concentration | p. 11 | UpdateConcentration(), Gas::AirO2, Sample::o2_permille |
API guide, Configuration |
| Stop measurement | SFM4300 D1: 4.3.4 Stop continuous measurement | p. 12 | Stop(), State::Measuring, State::StateUnknown |
API guide, Architecture, Contract |
| Averaging | SFM4300 D1: 3.1 Measurement modes; 4.3.5 Configure averaging | p. 12 | SetAveraging(), Config::averaging, Sample::smoothed |
API guide, Configuration |
| Scale factor, offset and unit | SFM4300 D1: 4.3.6 Read scale factor, offset and flow unit; 4.5 Unit conversion | p. 13 | ReadScaling(), ScalingInfo, Sample::flow_ml_per_min |
API guide, Configuration, Contract |
| Soft reset (general call) | SFM4300 D1: 4.3.7 Soft reset | p. 13 | SoftReset(), WriteGeneralCall(), Config::allow_general_call_reset |
API guide, Contract, Porting, Hardware |
| Sleep and wake-up | SFM4300 D1: 4.3.8 Entering and exiting sleep mode | p. 14 | Sleep(), WakeUp(), State::Asleep |
API guide, Architecture |
| CRC-8 of every word | SFM4300 D1: 4.4 Checksum calculation, Table 14 | p. 15 | reg::Crc8(), Error::Crc, reg::CrcVectorsHold() |
Contract, Testing and CI, Troubleshooting |
| Product identifier and serial number | SFM4300 D1: 4.3.9 Read product identifier, Table 13 | p. 15 | BeginInitialize(), GetProductInfo(), Error::DeviceIdMismatch |
Quickstart, API guide, Configuration, Troubleshooting |
| Fluidics, fittings and outline | SFM4300 D1: 5 Fluidic specifications; 6 Package outline | p. 17 | - | Hardware |
| Sfx6000 frame, start codes and addresses (preliminary) | SFx6000 I2C: I2C interface reference (preliminary) | p. 6 | Family::Sfx6000, Variant::SFM6000D_50, Capabilities::preliminary |
API guide, Configuration, Contract, Troubleshooting |
| Legacy command set with pointer registers (Sf05) | SFM-05: I2C functional description (commands, EEPROM, serial, CRC) | p. 4 | Family::Sf05, ProductInfo::serial, reg::kSf05Serial |
API guide, Architecture, Contract, Troubleshooting |
Feature coverage¶
Every row is a feature of the sensors' interface. "Yes" means the driver offers it and the host tests exercise it against the sensor model; the model is independent of the driver but written from the same documents. Nothing here has run on a sensor.
| Feature | SF06 | SF06 gen 2 | SF06 dual | SF05 | SF04 | SFx6000 [P] |
|---|---|---|---|---|---|---|
| Identify the part | product identifier | product identifier | product identifier | serial number (no part check) | no identifier | product identifier |
| Gases | air, O2; N2O, CO2 (SFM4300-20); heliox (SFM3013-CLM) | air | air (two channels) | air and N2; O2 on the SFM3000 only (the SFM4200 gives an oxygen correction and the SFM3200 refers to an application note: not applied, see K-O2-SF05) | one per variant | O2, air, CO2, N2O, Ar |
| Mixtures | air-O2 (and more on the SFM4300-20, CLM) | no | no | no | no | air-O2 |
UpdateConcentration() |
yes | no | no | no | no | yes |
Averaging (SetAveraging()) |
N = 0 to 128 | N = 0 to 128 and filter | N = 0 to 128 and filter | no | no | no |
| Scale, offset, unit read and checked | yes | yes | no (table only) | yes (EEPROM) | yes (EEPROM) | yes |
| Temperature | in the frame | in the frame | no | ReadTemperature() (SFM3000, -AW, SFM4200) |
no | ReadTemperature() |
| Status word checks | yes | yes | no | no | no | yes (concentration unchecked) |
| Sleep and wake | yes | yes | no | no | no | no |
| Soft reset | general call | general call | general call | command (SFM4200: /RES pin) | command | general call |
Hard reset (HardReset()) |
no | no | supply (port hook) | supply (port hook); /RES on SFM4200 | no | no |
| Clock-stretching read | no | no | no | no | required | no |
| Two flows in one frame, 24 bits | no | no | yes | no | no | no |
| Device-loss detection (silence past the datasheet bound, then the stop command) | yes | yes | yes | no (error limit instead) | no | yes |
Error limit, then StateUnknown and HardReset() by the caller |
- | - | - | yes | - | - |
The general-call reset needs Config::allow_general_call_reset and the port hook WriteGeneralCall(); without both
SoftReset() returns Unsupported.
Not in the driver¶
| Feature | Why |
|---|---|
| SFC controller products (valve, set-point) | Owner decision: out of scope; the flow commands are shared, the control commands are not implemented |
| SFM3020 and SFM3100 | Analog interface |
| SFM5300, SFM5330, SFM5400, SFM5460, SFM5500, SFC5xxx | RS485, SHDLC, DeviceNet or IO-Link; no I2C |
| SHDLC on the SFM6000 | A different bus; the I2C reference is the driver's source |
| Calibration-data and heater commands beyond those listed | Not documented in the sources read |
| An interrupt-driven mode that owns the IRQn pin | The driver does not own pins; the pin is the application's |
The SFM3003, SFM3013, SFM3019, SFM3119 and SFM3003-CET datasheets repeat the SFM4300 command set and differ in addresses, scale factors and timing; the SFM3003-CET adds only the in-channel gas temperature in the same frame layout.
Reviewed, not supported (evidence)¶
Twelve of the 32 documents describe other products or interfaces; they are in the register so that "not supported"
rests on a document: sfm3020, sfm3100, sfm5300_2016, sfx53x0, sfx5400, sfx5460, sfx5500, sfx5_shdlc,
sfx6000_shdlc, sfm_eng_guide (engineering background) and the two CRC and I2C application notes that apply to the