Troubleshooting

Find the cause of an error or a wrong reading. The first line of every answer is the error name that the call returned (sfm::ToString(error) gives it for logs).

Datasheet sections this page relies on
  • IRQn and start-up behaviour: Sensirion AG SFM4300 D1 1.0, 3.2 Sensor start-up and warm-up; 3.3 IRQn, p. 8
  • Start measurement, results and the NACK of a read without data: Sensirion AG SFM4300 D1 1.0, 4.3.1 Start continuous measurement, p. 10
  • Status word: Sensirion AG SFM4300 D1 1.0, 4.3.2 Read measurement results, Table 5, p. 11
  • CRC-8 of every word: Sensirion AG SFM4300 D1 1.0, 4.4 Checksum calculation, Table 14, p. 15
  • Product identifier and serial number: Sensirion AG SFM4300 D1 1.0, 4.3.9 Read product identifier, Table 13, p. 15
  • Sfx6000 frame, start codes and addresses (preliminary): Sensirion AG SFx6000 I2C 1.1, I2C interface reference (preliminary), p. 6
  • Legacy command set with pointer registers (Sf05): Sensirion AG SFM-05 not printed (vendor page: 10/2026), I2C functional description (commands, EEPROM, serial, CRC), p. 4

All of them, with the driver API for each: Datasheet map.

By error

Error Where you see it What to check
NoDevice Initialisation, a poll, a write, a read after a long silence Nothing acknowledged the address. Check the supply (see the variant table: the SF05 parts need 5 V), SDA and SCL wiring and pull-ups, and the address: the SFM4300 has 0x2A to 0x2D by its ADDR pin, the other parts a fixed one. A sleeping sensor does not answer until woken
DeviceIdMismatch Initialisation A sensor answered but it is not the configured part. Check Config::variant and address; with Variant::Auto check that the part is in the configured family. A prototype (revision 8x) needs accept_prototype
Bus Any write A write hook returned false: a NACK on a write, or a bus error. Check the pull-ups, the clock against the part's limit, and that the port sends a STOP
Timeout Wake, reset, a stretched read The sensor did not answer in time. After Sleep() the wake can take up to 20 ms; a reset up to 100 ms on the SF05 parts. A sensor held in reset, or SCL held low, stays silent
NoData ReadSample() Not a fault: no new sample yet. The SF06 parts make one every 0.5 ms; after an averaging of N, every N x 0.5 ms. Ask again later. After a silence longer than the datasheet bound the driver asks with a real command and returns DeviceReset or NoDevice
DeviceReset ReadSample() No new sample for longer than the datasheet bound, and the sensor acknowledged the command the driver then sent: it was reset (a power dip, a general call from another device, a brown-out) or is now stopped. The driver is Idle; start again. With IRQn wired, low after a start also means reset
Crc A read A word failed its CRC and the whole transfer was dropped. Check SCL speed, wire length, noise and the pull-ups. A line stuck low or high gives this, not a value
InvalidReply A read The CRC is valid but the reply cannot be sent by the sensor, for example an all-zero identifier. Check for a stuck bus or another device at the address
StatusMismatch ReadSample() The status word does not say the measurement the driver started is running (wrong gas index or concentration). Stop and start again; on a bus with another master, check that nothing else sent a start command
Stale ReadSample() The sample is inside the warm-up time (require_settled), smoothed (reject_smoothed) or older than max_age_us. Not a fault; the policy is yours
OutOfRange ReadSample() The flow is beyond the part's span plus the margin. Check the installation and the gas; RangePolicy::Flag returns the value flagged
DeviceFault StartMeasurement(), ReadScaling() The scale, offset or unit the sensor reports differs from the datasheet. A recalibrated part, a wrong variant, or a corrupted sensor. scaling_check chooses how strict to be
StateUnknown After a failed command, or at the SF05 error limit The driver cannot know the sensor's mode. Call EnsureInitialized() (or SoftReset() first)
Busy Any call An operation is pending (finish it with Poll() or AwaitPending()), or the call is illegal in this mode: configuration only when idle, a read only while measuring, a wake only when asleep
NotInitialized Any call EnsureInitialized() has not succeeded, or a reset or Deinitialize() ran since
Unsupported A call The part, the port or the family mask does not offer the call: SoftReset() needs allow_general_call_reset and the port's WriteGeneralCall(); HardReset() needs PowerCycle() or SetResetPin(); Sleep() is for the SF06 parts and SFM3304
InvalidArgument A call, initialisation A Config field or argument is out of range; Validate(config) says so without touching the bus. Typical: an address outside the part's set, Auto with TimingPolicy::Datasheet, averaging on a part without it, an oxygen fraction above 1000

By symptom

Symptom Likely cause
A flow that is always 0 on an SFM4200 or SFM4100 An all-zero frame has a valid CRC there (I9 exception); check the bus level and GetStats()
Flow reads too high or too low by a fixed factor The wrong gas: StartMeasurement(gas) selects the calibration. A mixture needs its oxygen fraction (UpdateConcentration())
The first samples are off The warm-up is 30 ms after a start (50 ms on the SFM3304 and SFM3505); use Sample::settled
Readings stop after a while on an SF05 part A lock-up after an aborted read: the port must never abort a read early. At the error limit the driver is StateUnknown; call HardReset() (the port's PowerCycle(), which must hold SDA and SCL low while the supply is off) and initialise again
Another device on the bus resets SoftReset() is a general call and resets every device that listens. Leave allow_general_call_reset off when the bus is shared
Two sensors at one address 0x2E is the SFM3019, SFM3304 and SFM3505; 0x40 is every SF05 part. Use the SFM4300's ADDR pin or separate buses
Nothing on an SFM4100 It needs a clock-stretching read: the port needs ReadStretch() and a controller that tolerates a stretching slave
An SFM3304 not found after the MCU reset It was asleep. Send the general call reset or cycle the supply, then initialise

Counters

GetStats() returns crc_errors, no_data, bus_errors, device_resets, consecutive_errors and power_cycles. A rising crc_errors with a clean bus_errors is a signal-integrity problem; a rising device_resets is a supply problem.