| struct | AddressScan |
| | The Modbus addresses that answered a scan: bit n of byte n/8 is address n. More...
|
| |
| struct | Averages |
| | Result of ReadAverages(): the current or most recent measurement window. More...
|
| |
| class | Basis2 |
| | Driver for one instrument on a Bus. More...
|
| |
| class | Bus |
| | The transaction machine of one serial line, shared by every instrument on it. More...
|
| |
| struct | Config |
| | The driver's configuration. Every default is the instrument's factory value or a documented driver choice; fields that exist because the manual is silent are described in docs/configuration.md. More...
|
| |
| struct | ConfirmAbandonProtocol2 |
| | Confirms SwitchToProtocol1(): the instrument leaves the protocol this driver speaks. More...
|
| |
| struct | ConfirmAddressChange |
| | Confirms SetModbusAddress(): the Modbus address changes at once. More...
|
| |
| struct | ConfirmBaudChange |
| | Confirms ChangeBaud(): the reply arrives at the new speed and the port must follow. More...
|
| |
| struct | ConfirmFactoryRestore |
| | Confirms FactoryRestore(): unit ID, baud rate, Modbus address and protocol revert; power cycle. More...
|
| |
| struct | ConfirmReturnToProtocol2 |
| | Guard token of SwitchToProtocol2. More...
|
| |
| struct | ConfirmUnitIdChange |
| | Confirms ChangeUnitId(): the instrument answers to the new ID afterwards; connect one instrument only. More...
|
| |
| struct | Decimal |
| | An exact decimal number as the instrument printed it: +0021513.0 is {215130, 1}. More...
|
| |
| struct | Frame |
| | One data frame (A) or a DV selection; present says which fields were read. More...
|
| |
| struct | FullScale |
| | One reading of FPF: a value and its unit text. More...
|
| |
| struct | Gains |
| | Closed-loop gains (LCG, Modbus 519 and 520). More...
|
| |
| struct | GasEntry |
| | One entry of the gas table of firmware 3.0.5 (Modbus 81 to 140, M p.26 to 27). More...
|
| |
| struct | GasId |
| | A gas number. Not a closed enumeration: the V3.0.5 gas table allows numbers up to 255 (A5). More...
|
| |
| struct | GasList |
| | Result of ReadGasList(): the parsed pairs, or the raw text when the grammar did not match (A5). More...
|
| |
| struct | GasTable |
| | The whole gas table: twelve entries. More...
|
| |
| struct | Identity |
| | What Begin() learned: read-only knowledge of the instrument. More...
|
| |
| struct | PreviousMeasurement |
| | Result of ReadPreviousMeasurement() (Modbus only, registers 4210 to 4215). More...
|
| |
| struct | Ramp |
| | Result of ReadRamp(): rate and time unit as the instrument reported them. More...
|
| |
| struct | Ranges |
| | Result of ReadRanges(). More...
|
| |
| struct | Reply |
| | A reply as received: one ASCII line without its CR, or one Modbus frame (raw bytes as chars). More...
|
| |
| class | Request |
| | One validated request. Only the factories of the grammar headers make one that can be sent. More...
|
| |
| struct | Result |
| | The value of a call, or why there is none. More...
|
| |
| struct | Result< void > |
| | The result of a call that returns no value. More...
|
| |
| struct | StatusFlags |
| | Status flags of a data frame or of Modbus register 2101 (M p.10, p.21). More...
|
| |
| struct | Text |
| | A short text value (serial number, firmware, unit label), NUL terminated. More...
|
| |
| struct | Timing |
| | What the bus needs to know about the instrument it talks to, per transaction. More...
|
| |
| class | UartInterface |
| | CRTP base of the platform serial port. More...
|
| |
| struct | UnitIdScan |
| | The unit IDs that answered a scan: bit 0 is A. More...
|
| |
|
| enum class | Command : uint8_t {
Command::Poll = 0
, Command::ReadData
, Command::SetSetpoint
, Command::ReadSetpointSource
,
Command::SetSetpointSource
, Command::ReadRamp
, Command::SetRamp
, Command::DisableRamp
,
Command::ReadWatchdog
, Command::SetWatchdog
, Command::HoldValve
, Command::ResumeControl
,
Command::ReadGains
, Command::SetGains
, Command::ReadReferenceTemperature
, Command::SetReferenceTemperature
,
Command::ReadAveraging
, Command::SetAveraging
, Command::ReadTrigger
, Command::SetTrigger
,
Command::StartMeasurement
, Command::ReadAverages
, Command::ReadRanges
, Command::Tare
,
Command::ReadAutotare
, Command::SetAutotare
, Command::ReadGas
, Command::ReadGasList
,
Command::SetGas
, Command::ResetTotal
, Command::SetBatch
, Command::DisableBatch
,
Command::ReadTotalizerLimit
, Command::SetTotalizerLimit
, Command::ReadFullScale
, Command::ChangeUnitId
,
Command::ReadSerialNumber
, Command::ReadFirmware
, Command::ReadModbusAddress
, Command::SetModbusAddress
,
Command::ReadBaud
, Command::ChangeBaud
, Command::ReadProtocol
, Command::SwitchToProtocol1
,
Command::SwitchToProtocol2
, Command::FactoryRestore
, Command::Modbus
} |
| | The 46 ASCII commands of M pp.10 to 17 (rows C01 to C46 of the coverage map), plus the Modbus kind. More...
|
| |
| enum class | Error : uint8_t {
Error::Ok = 0
, Error::NotInitialized
, Error::InvalidArgument = 2
, Error::Bus = 3
,
Error::Timeout = 4
, Error::DeviceIdMismatch
, Error::Crc = 6
, Error::Busy
,
Error::DeviceFault = 8
, Error::Unsupported
, Error::BadReply = 16
, Error::Rejected = 17
,
Error::OutOfRange = 18
, Error::Overflow = 19
, Error::VerifyMismatch = 20
, Error::FirmwareTooOld = 21
,
Error::StateUnknown
, Error::RateLimited
, Error::GuardNotConfirmed = 24
} |
| | Why a call failed. Every call of the driver returns one of these in its Result. More...
|
| |
| enum class | LogLevel : uint8_t { LogLevel::Error = 0
, LogLevel::Warning
, LogLevel::Info
, LogLevel::Debug
} |
| | Severity passed to the optional Log() hook. More...
|
| |
| enum class | Protocol : uint8_t { Protocol::Ascii = 0
, Protocol::ModbusRtu = 1
} |
| | The command protocol one Basis2 object speaks. More...
|
| |
| enum class | DeviceKind : uint8_t { DeviceKind::Auto = 0
, DeviceKind::Meter = 1
, DeviceKind::Controller = 2
} |
| | Whether the instrument is a meter (B series) or a controller (BC series). More...
|
| |
| enum class | FrameLayout : uint8_t { FrameLayout::Auto = 0
, FrameLayout::Meter = 1
, FrameLayout::Controller = 2
} |
| | Which data frame layout is accepted (assumption A2: a meter's frame is not shown in the manual). More...
|
| |
| enum class | UnknownFlagPolicy : uint8_t { UnknownFlagPolicy::Report = 0
, UnknownFlagPolicy::Reject = 1
} |
| | What happens with a flag token the manual does not list. More...
|
| |
| enum class | EchoMode : uint8_t { EchoMode::None = 0
, EchoMode::Skip = 1
} |
| | Whether a half-duplex wiring that hears its own transmission is handled. More...
|
| |
| enum class | VerifyPolicy : uint8_t { VerifyPolicy::Off = 0
, VerifyPolicy::WhenNoRamp
} |
| | Whether a setpoint is read back after it was set. More...
|
| |
| enum class | FirmwareGate : uint8_t { FirmwareGate::Enforce
, FirmwareGate::Ignore = 1
} |
| | Whether the firmware gates of the register and command tables are enforced. More...
|
| |
| enum class | SetpointSource : uint8_t { SetpointSource::Analog = 0
, SetpointSource::SavedDigital
, SetpointSource::UnsavedDigital = 2
} |
| | Setpoint source (LSS, Modbus 516; M p.11 to 12 and p.19). More...
|
| |
| enum class | TimeUnit : uint8_t { TimeUnit::Millisecond = 3
, TimeUnit::Second = 4
, TimeUnit::Minute = 5
} |
| | Time unit of the ramp command (SR r t, M p.12). More...
|
| |
| enum class | TotalizerLimit : uint8_t { TotalizerLimit::HoldNoError = 0
, TotalizerLimit::ResetNoError = 1
, TotalizerLimit::HoldWithError = 2
, TotalizerLimit::ResetWithError = 3
} |
| | Totalizer limit behaviour (TC, Modbus 54; M p.16). More...
|
| |
| enum class | FullScaleItem : uint8_t { FullScaleItem::Flow = 0
, FullScaleItem::TotalVolume = 1
, FullScaleItem::MaxTemperature = 2
} |
| | The item read by FPF (M p.16). More...
|
| |
| enum class | DataItem : uint8_t {
DataItem::Flow = 1
, DataItem::Setpoint = 2
, DataItem::Temperature = 4
, DataItem::ValveDrive = 8
,
DataItem::Gas = 16
, DataItem::TotalVolume = 32
, DataItem::BatchRemaining = 64
, DataItem::Status = 128
} |
| | Items of DV (a bit mask, M p.11). More...
|
| |
| enum class | TriggerMode : uint8_t { TriggerMode::OnSetpointChange = 1
, TriggerMode::OnHoldChange = 2
, TriggerMode::OnAverageRead = 4
} |
| | Measurement trigger bits (MT, Modbus 4200; M p.13 to 14). More...
|
| |
| enum class | Baud : uint8_t {
Baud::B4800 = 0
, Baud::B9600 = 1
, Baud::B19200 = 2
, Baud::B38400 = 3
,
Baud::B57600 = 4
, Baud::B115200 = 5
} |
| | Serial speeds the instrument offers (NCB, Modbus 21; M p.17, p.25). More...
|
| |
| enum class | CommandProtocol : uint8_t { CommandProtocol::BasisProtocol2 = 0
, CommandProtocol::BasisProtocol1 = 1
} |
| | Command protocol selector (P, Modbus 56). More...
|
| |
| enum class | FlowUnit : uint8_t {
Sccm = 0
, Nccm = 1
, Slpm = 2
, Nlpm = 3
,
SmLs = 4
, NmLs = 5
, SmLm = 6
, NmLm = 7
,
SLh = 8
, NLh = 9
, Sccs = 10
, Nccs = 11
,
Sm3h = 12
, Nm3h = 13
, Sm3d = 14
, Nm3d = 15
,
Scim = 16
, Scfm = 17
, Scfh = 18
, Scfd = 19
,
FlowUnit::Other = 20
} |
| | Flow units (Modbus register 49, M p.26; M p.31). More...
|
| |
| enum class | PollState : uint8_t { Idle = 0
, Pending = 1
, Done = 2
, Failed = 3
} |
| | State of a transaction as seen by Bus::Poll(). More...
|
| |
| enum class | BusState : uint8_t { Idle = 0
, Gap = 1
, AwaitReply = 2
, Settling = 3
} |
| | Internal phase of the bus. More...
|
| |
|
| constexpr Config | KconfigDefaults () noexcept |
| | The configuration menuconfig selected, on top of the datasheet defaults.
|
| |
| constexpr const char * | ToString (Error error) noexcept |
| | The name of an error, for logs.
|
| |
| constexpr uint32_t | GetDriverVersionMajor () noexcept |
| | Major version of the driver, from the generated alicat_basis2_version.h.
|
| |
| constexpr uint32_t | GetDriverVersionMinor () noexcept |
| | Minor version of the driver.
|
| |
| constexpr uint32_t | GetDriverVersionPatch () noexcept |
| | Patch version of the driver.
|
| |
| constexpr const char * | GetDriverVersion () noexcept |
| | The version as text, such as "0.1.0".
|
| |
| constexpr Error | Validate (const Config &c) noexcept |
| | Check a configuration against the documented ranges.
|
| |
| constexpr const char * | GasName (uint8_t number) noexcept |
| | The name of a gas of the nine of M p.15.
|
| |
| constexpr uint8_t | operator| (DataItem a, DataItem b) noexcept |
| | Union of DataItem bits.
|
| |
| constexpr uint8_t | operator| (uint8_t a, DataItem b) noexcept |
| | Union of a mask and an item.
|
| |
| constexpr StatusFlags | StatusFromBits (uint32_t bits) noexcept |
| | Decode status bits (Modbus register 2101 or an integer DV status): +1 MOV, +2 TOV, +4 OVR, +8 HLD, +16 VTM.
|
| |
|
|
constexpr std::size_t | kMaxRequest = 48 |
| | Longest request the driver builds, in bytes.
|
| |
|
constexpr uint8_t | kDecimalsUnset = 0xFF |
| | "Not set" marker of Config::modbus_flow_decimals: Begin() fails in Modbus mode while it is.
|
| |
|
constexpr std::size_t | kMaxLine = 256 |
| | Largest line or frame the driver buffers: the gas list (GS *) reply can be this long.
|
| |
|
constexpr std::size_t | kTextMax = 24 |
| | Maximum characters kept of a text reply.
|
| |
|
constexpr uint8_t | kFlagCount = 5 |
| | Position of the unit ID in a StatusFlags text: helper used by tests.
|
| |
|
What the instrument needs from the serial line and its supply (manual pp. 6, 10 and 17). A port and a board configuration check their settings against these with static_assert (handbook KB-CPP-123).
|
|
constexpr uint32_t | kPortDefaultBaud = 38'400 |
| | The baud rate after a factory restore, 8 data bits, no parity, 1 stop bit (M p.10).
|
| |
|
constexpr uint32_t | kMinBaudRate = 4'800 |
| | The lowest baud rate the instrument offers (M p.17).
|
| |
|
constexpr uint32_t | kMaxBaudRate = 115'200 |
| | The highest baud rate the instrument offers, firmware 2.2.2 or later (M p.17).
|
| |
| constexpr uint32_t | kSilentBitTimes = 35 |
| |
| constexpr uint32_t | kTurnaroundUs = 50'000 |
| |
|
constexpr uint32_t | kSupplyMinMv = 12'000 |
| | Lowest supply, millivolts, on pin 4 only (M p.6).
|
| |
|
constexpr uint32_t | kSupplyMaxMv = 24'000 |
| | Highest supply, millivolts (M p.6).
|
| |
The ALICAT_BASIS2 driver: everything it declares lives here; internals in detail.