BASIS 2 Driver 0.1.0
API reference
|
The driver's configuration. Every default is the instrument's factory value or a documented driver choice; fields that exist because the manual is silent are described in docs/configuration.md.
More...
#include <alicat_basis2_types.hpp>
Public Attributes | |
| char | unit_id {'A'} |
ASCII unit ID, A to Z (M p.10). | |
| Protocol | protocol {Protocol::Ascii} |
| The protocol this object speaks. | |
| uint8_t | modbus_address {1} |
| Modbus address 1 to 247. | |
| uint32_t | baud {38400} |
| What the port runs at; sizes every timeout. | |
| DeviceKind | kind {DeviceKind::Auto} |
Meter or controller; Auto learns it at Begin(). | |
| uint16_t | processing_ms {50} |
| Instrument turnaround added to every reply wait. | |
| uint16_t | slow_command_ms {1000} |
| Reply wait of commands that may write non-volatile memory. | |
| uint16_t | restore_holdoff_ms {3000} |
| Quiet time after FACTORY RESTORE. | |
| uint8_t | read_retries {1} |
| Retries of a read after a failed transaction; writes never repeat. | |
| FrameLayout | frame_layout {FrameLayout::Auto} |
| Which data frame layout is accepted. | |
| UnknownFlagPolicy | unknown_flags {UnknownFlagPolicy::Report} |
| What happens with a status flag the manual does not list. | |
| EchoMode | echo {EchoMode::None} |
| Whether a first line equal to the command is dropped (half-duplex wiring that hears itself). | |
| uint16_t | flow_margin_permille {1100} |
| Plausibility: the flow must not exceed the full scale times this over 1000. | |
| int16_t | temperature_min_c {-50} |
| Plausibility lower limit of the temperature, C. | |
| int16_t | temperature_max_c {150} |
| Plausibility upper limit of the temperature, C. | |
| uint8_t | setpoint_decimals {3} |
| Digits sent after the point in a setpoint. | |
| VerifyPolicy | verify_setpoint {VerifyPolicy::WhenNoRamp} |
| Whether a setpoint is read back. | |
| uint16_t | verify_tolerance_permille {5} |
| Read-back tolerance in permille of the full scale. | |
| uint32_t | saved_setpoint_min_interval_ms {180000} |
| Minimum time between setpoints while the source is saved. | |
| FirmwareGate | firmware_gate {FirmwareGate::Enforce} |
| Whether the firmware gates of the register and command tables are enforced. | |
| uint16_t | assume_firmware {0} |
0 unknown; else 256a + 16b + c, used when the instrument gives no firmware. | |
| uint8_t | modbus_flow_decimals {kDecimalsUnset} |
| Required in Modbus mode: 0 to 6 (the decimals are not readable over Modbus). | |
| uint32_t | full_scale_override_milli {0} |
0 learns the full scale at Begin(); else flow units times 1000. | |
The driver's configuration. Every default is the instrument's factory value or a documented driver choice; fields that exist because the manual is silent are described in docs/configuration.md.