BASIS 2 Driver 0.1.0
API reference
Loading...
Searching...
No Matches
Classes | Enumerations | Variables
Requests and replies

The command set and the timing formulas. More...

Classes

struct  alicat_basis2::Reply
 A reply as received: one ASCII line without its CR, or one Modbus frame (raw bytes as chars). More...
 
class  alicat_basis2::Request
 One validated request. Only the factories of the grammar headers make one that can be sent. More...
 

Enumerations

enum class  alicat_basis2::Command : uint8_t {
  alicat_basis2::Command::Poll = 0 , alicat_basis2::Command::ReadData , alicat_basis2::Command::SetSetpoint , alicat_basis2::Command::ReadSetpointSource ,
  alicat_basis2::Command::SetSetpointSource , alicat_basis2::Command::ReadRamp , alicat_basis2::Command::SetRamp , alicat_basis2::Command::DisableRamp ,
  alicat_basis2::Command::ReadWatchdog , alicat_basis2::Command::SetWatchdog , alicat_basis2::Command::HoldValve , alicat_basis2::Command::ResumeControl ,
  alicat_basis2::Command::ReadGains , alicat_basis2::Command::SetGains , alicat_basis2::Command::ReadReferenceTemperature , alicat_basis2::Command::SetReferenceTemperature ,
  alicat_basis2::Command::ReadAveraging , alicat_basis2::Command::SetAveraging , alicat_basis2::Command::ReadTrigger , alicat_basis2::Command::SetTrigger ,
  alicat_basis2::Command::StartMeasurement , alicat_basis2::Command::ReadAverages , alicat_basis2::Command::ReadRanges , alicat_basis2::Command::Tare ,
  alicat_basis2::Command::ReadAutotare , alicat_basis2::Command::SetAutotare , alicat_basis2::Command::ReadGas , alicat_basis2::Command::ReadGasList ,
  alicat_basis2::Command::SetGas , alicat_basis2::Command::ResetTotal , alicat_basis2::Command::SetBatch , alicat_basis2::Command::DisableBatch ,
  alicat_basis2::Command::ReadTotalizerLimit , alicat_basis2::Command::SetTotalizerLimit , alicat_basis2::Command::ReadFullScale , alicat_basis2::Command::ChangeUnitId ,
  alicat_basis2::Command::ReadSerialNumber , alicat_basis2::Command::ReadFirmware , alicat_basis2::Command::ReadModbusAddress , alicat_basis2::Command::SetModbusAddress ,
  alicat_basis2::Command::ReadBaud , alicat_basis2::Command::ChangeBaud , alicat_basis2::Command::ReadProtocol , alicat_basis2::Command::SwitchToProtocol1 ,
  alicat_basis2::Command::SwitchToProtocol2 , alicat_basis2::Command::FactoryRestore , alicat_basis2::Command::Modbus
}
 The 46 ASCII commands of M pp.10 to 17 (rows C01 to C46 of the coverage map), plus the Modbus kind. More...
 

Variables

constexpr std::size_t alicat_basis2::kMaxRequest = 48
 Longest request the driver builds, in bytes.
 

Detailed Description

The command set and the timing formulas.

Enumeration Type Documentation

◆ Command

enum class alicat_basis2::Command : uint8_t
strong

The 46 ASCII commands of M pp.10 to 17 (rows C01 to C46 of the coverage map), plus the Modbus kind.

Enumerator
Poll 

C01 a.

ReadData 

C02 aDV n.

SetSetpoint 

C03 aS f.

ReadSetpointSource 

C04 aLSS.

SetSetpointSource 

C05 aLSS x.

ReadRamp 

C06 aSR.

SetRamp 

C07 aSR r t.

DisableRamp 

C08 aSR 0.

ReadWatchdog 

C09 aWD.

SetWatchdog 

C10 aWD t.

HoldValve 

C11 aHPUR p.

ResumeControl 

C12 aC.

ReadGains 

C13 aLCG.

SetGains 

C14 aLCG p i.

ReadReferenceTemperature 

C15 aRT.

SetReferenceTemperature 

C16 aRT t.

ReadAveraging 

C17 aDCA.

SetAveraging 

C18 aDCA t.

ReadTrigger 

C19 aMT.

SetTrigger 

C20 aMT m.

StartMeasurement 

C21 aDVAS t.

ReadAverages 

C22 aDVAA.

ReadRanges 

C23 aDVAR.

Tare 

C24 aV d.

ReadAutotare 

C25 aZCA.

SetAutotare 

C26 aZCA 1 or aZCA 0.

ReadGas 

C27 aGS.

ReadGasList 

C28 aGS *.

SetGas 

C29 aGS n.

ResetTotal 

C30 aT.

SetBatch 

C31 aTB v.

DisableBatch 

C32 aTB 0.

ReadTotalizerLimit 

C33 aTC.

SetTotalizerLimit 

C34 aTC m.

ReadFullScale 

C35 aFPF v.

ChangeUnitId 

C36 a@=b (guarded).

ReadSerialNumber 

C37 aSN.

ReadFirmware 

C38 aVE.

ReadModbusAddress 

C39 aMA.

SetModbusAddress 

C40 aMA n (guarded).

ReadBaud 

C41 aNCB.

ChangeBaud 

C42 aNCB b (guarded).

ReadProtocol 

C43 aP.

SwitchToProtocol1 

C44 aP 1 (guarded).

SwitchToProtocol2 

C45 aP2.

FactoryRestore 

C46 aFACTORY RESTORE (guarded).

Modbus 

A Modbus RTU frame (function 3, 6 or 16).