Troubleshooting

Symptoms, causes and fixes, cheapest check first. Nothing here has been confirmed on an instrument; the causes are the ones the manual and the line rules point to.

Datasheet sections this page relies on

From Alicat Scientific, Inc. DOC-MANUAL-BASIS2 Rev 4:

  • Unit ID: Unit ID, p. 16
  • Modbus address, baud rate, command protocol: Modbus Address; Serial Baud Rate; Command Protocol, p. 17
  • Troubleshooting: Troubleshooting, p. 29

All of them, with the driver API for each: Datasheet map.

No reply

Begin() or any read returns Timeout: no complete line arrived within the reply wait (76 ms at 38400 baud).

Check Fix
The supply is on pin 4, 12 to 24 V, and the LED is green Power the instrument; never connect power to another pin
TX and RX Pin 5 is the instrument's receive (or RS-485 B), pin 6 its transmit (or A): cross them for RS-232
RS-485 A and B Swap the pair: a swapped pair gives garbage or silence, and the labels differ between vendors
RS-485 direction The port must release the transmitter after the last byte has left the pin (Porting)
Baud rate Config::baud must equal the instrument's (factory 38400); try the other five rates one at a time
Unit ID Config::unit_id must equal the instrument's (factory A); ScanUnitIds() finds it
Protocol The manual does not say whether ASCII and Modbus work at once on one line; try Config::protocol the other way
Levels A TTL UART cannot talk to an RS-232 or RS-485 instrument without a level shifter or a transceiver

Garbage or BadReply

A line arrived but does not parse. The usual causes are a wrong baud rate (bytes that are not printable ASCII), a half duplex wiring that hears its own transmission (set Config::echo = EchoMode::Skip), or an instrument speaking BASIS Protocol 1 (the older protocol, which the driver does not speak; see below). A reply that is longer than the 256-byte buffer without a carriage return is Overflow: the baud rate is almost always wrong.

DeviceIdMismatch

An instrument answered with another unit ID than the one addressed. Two instruments with one ID answer together and corrupt each other's data (manual p. 16): connect them one at a time and give each its own ID, with ChangeUnitId() on a line that has one instrument (Safety notes).

Crc (Modbus)

Single corrupted frames point at noise or a missing termination. Every frame fails: a baud-rate or parity mismatch (the driver assumes no parity), or ASCII wired to a Modbus instrument. A frame that fails the CRC is discarded whole and retried (reads only).

The flow is ten times wrong over Modbus

Config::modbus_flow_decimals does not match the instrument's factory decimals. Over ASCII the decimals are printed in the frame; read one frame over ASCII, count the decimals and set the same number.

Rejected

The instrument answered ? or a Modbus exception. A controller-only command on a meter, a value outside the instrument's range, or P 1 with a flow unit that BASIS Protocol 1 does not have (only SCCM, SLPM, NCCM and NLPM are allowed). The driver learns the kind at Begin() and returns Unsupported for a controller command on a meter without sending.

The setpoint is ignored

The source is Analog: a digital setpoint command is not accepted (ReadSetpointSource(); set it with SetSetpointSource(UnsavedDigital)). Or a ramp limits the change (VerifyMismatch after a ramp is expected; the verify is skipped while a ramp is active). Or the valve is held (HLD flag): call ResumeControl().

RateLimited

The source is SavedDigital and the last setpoint was less than three minutes ago. The saved source writes non-volatile memory with every setpoint. Use UnsavedDigital.

VerifyMismatch

The value read back differs from the one set by more than Config::verify_tolerance_permille. The instrument accepts a setpoint up to 2.5 % over full scale and may round; a ramp or a clamp differs on purpose. Over Modbus, check the decimals.

StateUnknown

A guarded command may have changed the unit ID, the baud rate, the Modbus address or the protocol, and the driver could not confirm it. Call Recover(): it polls the configured ID, scans the unit IDs (or addresses) when the bus has one instrument, and relearns the instrument; a different instrument is VerifyMismatch. After a factory restore, power-cycle first. If the baud rate changed, set the port's speed to the new rate before Recover().

Stray lines and a streaming instrument

Bus::StrayLineCount() counts lines that arrive while no transaction waits; Bus::StreamingSuspected() becomes true when one begins with @. The driver never uses such a line and never sends a streaming command. Streaming mode is a feature of the vendor's classic line that is not in the BASIS 2 manual; if an instrument streams, switch it off with the vendor's software.

An instrument is stuck in BASIS Protocol 1

P 1 switches to the older protocol, whose grammar the driver does not implement. The manual says to return with P2, but a Protocol 1 instrument may not parse the line the driver sends. Use the vendor's software or a terminal to send P2 (no space) with the unit ID, or restore the factory settings.

The flow reads a few percent off

Tare with true zero flow on the pure process gas (Tare()), select the right gas (SetGas()), and check the reference temperature (ReadReferenceTemperature()): the standard flow is referred to it.

Wrong units

The driver prints and returns values in the instrument's own flow unit: GetIdentity().unit_label.chars.data(). Nothing converts.