BASIS 2 Driver

A C++20 driver for the Alicat BASIS 2 gas mass flow meters (B series) and controllers (BC series). It reads flow, temperature and totals, sets setpoints and valve holds, and speaks both the ASCII protocol and Modbus RTU behind one API, on any MCU.

Quickstart API guide Source on GitHub

The driver has no platform code. You give it a small class that moves bytes on a serial line and tells the time; it does the rest: the unit ID and silent-interval rules of the line, frame parsing, unit and decimal handling, Modbus framing and CRC, retries of reads, and the guards around the commands that can strand an instrument.

Gas mass flow meters and controllers BASIS 2 Alicat Scientific, B and BC series
Function
Mass flow, temperature and total of nine gases, 100 SCCM to 100 SLPM full scale; controllers add a setpoint, a ramp and a valve
Interface
RS-232 or RS-485 by model, 38400 baud 8N1 by default; ASCII protocol with unit IDs A to Z, or Modbus RTU
Protocols
ASCII "BASIS Protocol 2" (the default) and Modbus RTU, both complete, one API
Supply
12 to 24 V on pin 4 only; 210 mA for a controller, 12 mA for a meter
Connector
Six-pin JST GH: analog setpoint, analog output, ground, power, serial B and serial A

What you get

  • One object per instrument. Basis2<YourPort> sits on a Bus<YourPort> that owns the serial line, so up to 26 instruments share one line. See API guide.
  • Two protocols, one API. Config::protocol selects ASCII or Modbus RTU; the calls do not change. See Protocols.
  • Errors you can act on. Every call returns Result<T> with an Error such as Timeout, DeviceIdMismatch or VerifyMismatch. See Troubleshooting.
  • Guarded commands. The six commands that can leave an instrument unreachable need a token that names the consequence. See Safety notes.
  • A port of five hooks. Write, Read, FlushInput, NowMs and EnsureInitialized, plus optional DelayMs, SetBaud and Log. See Porting.
  • No heap, no exceptions, no RTTI. See Contract.
  • Tested without hardware. An independent model instrument with a byte-accurate wire, fault injection and four fuzz targets. See Testing and CI. Nothing has been run on an instrument yet.

Datasheet

Document Revision Get it
Datasheet manual: Operating Manual for Mass Flow Instruments, BASIS 2 Meters and Controllers (Alicat Scientific, Inc. DOC-MANUAL-BASIS2) Rev 4, 2025-05-23 PDF · vendor page
Datasheet specification sheet: Technical Data for BASIS 2.0 Mass Flow Meters and Controllers (Alicat Scientific, Inc. DOC-SPECS-BASIS) Rev 19, 2026-08-17 PDF · vendor page
Datasheet specification sheet: Technical Data for BASIS 2 (pages 1 to 3 of the combined manual and specification file) (Alicat Scientific, Inc. MANUAL+SPECS_BASIS2) Rev 16, 2024-03-01 PDF · vendor page
Datasheet manual: Alicat Modbus Manual (generic, written for the classic registers) (Alicat Scientific, Inc. DOC-MANUAL-MODBUS) Rev 7, 2024-11-01 PDF · vendor page
Datasheet application note: Alicat Serial Communications Primer (classic line, firmware 4v to 10v) (Alicat Scientific, Inc. Alicat-Serial-Primer) Rev 2, 2023-02-01 PDF · vendor page

Written against these revisions; the SHA-256 of each file is in docs/datasheets.yml. Where each part of the driver is in the datasheet: Datasheet map.

Where to go next

  • Quickstart


    Wire the instrument, write the port, read a flow in about 20 lines.

  • Hardware


    The connector, wiring for RS-232 and RS-485, power, LED and the size classes.

  • Safety notes


    Valve holds, the watchdog, saved setpoints, tare, and the guarded commands.

  • API guide


    Start up, read, set, scan and recover, task by task.

  • Porting


    Implement the port, RS-485 direction and baud switching.

  • API reference


    Every class and member, generated from the headers.