sfm::Config Struct Reference

Everything EnsureInitialized() applies. More...

#include <sfm_flow_meter_types.hpp>

Public Attributes

Variant variant {Variant::Auto}
 The part; Auto identifies it inside family.
 
Family family {Family::Sf06}
 Used with Auto only; never crosses families.
 
uint8_t address {0}
 7-bit address; 0 is the part's default.
 
uint16_t averaging {kAveragingDefault}
 N (0 average-until-read, 1..128) or the default.
 
bool filter_enabled {false}
 Send the filter argument with the start command.
 
uint16_t filter {kFilterDefault}
 The 16-bit signal-filter argument.
 
ScalingCheck scaling_check {ScalingCheck::Strict}
 See ScalingCheck.
 
RangePolicy range_policy {RangePolicy::Reject}
 See RangePolicy.
 
uint16_t range_margin_permille {20}
 Of the span, added at both ends before OutOfRange.
 
bool require_settled {false}
 A sample before the warm-up time is Stale (needs a clock).
 
bool reject_smoothed {false}
 A smoothed sample is Stale.
 
uint32_t max_age_us {0}
 
bool accept_prototype {false}
 A prototype product identifier is accepted.
 
bool allow_general_call_reset {false}
 
TimingPolicy timing {TimingPolicy::Conservative}
 See TimingPolicy.
 
uint8_t consecutive_error_limit
 SF05: errors in a row before a hard reset or StateUnknown.
 
bool restart_before_read {true}
 SF05: send the start command again after each read.
 
bool poll_with_identifier {false}
 
bool sf05_serial_low_word {true}
 

Detailed Description

Everything EnsureInitialized() applies.

Config{} is valid: automatic identification in the SF06 family, the datasheet reset behaviour, strict scaling checks and conservative timing. It has no operator==: nothing here is compared as a whole.

Member Data Documentation

◆ allow_general_call_reset

bool sfm::Config::allow_general_call_reset {false}

Permit the general-call reset (it resets every device that listens).

◆ consecutive_error_limit

uint8_t sfm::Config::consecutive_error_limit
Initial value:
{
5}

SF05: errors in a row before a hard reset or StateUnknown.

◆ max_age_us

uint32_t sfm::Config::max_age_us {0}

0 off; else a read later than this after the previous good one is Stale.

◆ poll_with_identifier

bool sfm::Config::poll_with_identifier {false}

SF06: idle polls send the identifier command, not stop (ambiguity K-STOPIDLE).

◆ sf05_serial_low_word

bool sfm::Config::sf05_serial_low_word {true}

SF05: also read the low serial word through 0x31AF (ambiguity K-SF05-SERIAL).